Overview
Specifications
Sample Data
NanoFrontier eLD w/ ECD main page
NanoFrontier LD main page
Contact Us


Product Number:NanoFrontier nLC

 
 

Sample Data

 

Gradient reproducibility at a flow rate of 50 nL/min (n=5)

[Conditions]

  • Flow rate: 50 nL/min
  • Solvents: A - Water; B - 80% CH3CN 50mg/mL
  • Gradient settings: 0 min (B=10%) ’! 40 min (B-100%); Detector UV 273 nm (31-nL cell, 1-nm slit)

An exampleof LC/MS measurement at a flow rate of 50 nL/min

[Conditions]

  • Sample: BSA
  • Sample amount: 1¼LL
  • Flow rate: 50 nL/min
  • Mobile phase: A) 0.1% formic acid in 2% acetonitrile; B) 0.1% formic acid in 98% acetonitrile
  • Gradient settings: A/B=98/2 (0 min) ’! 60/40 (60 min) ’!
    2/98 (60~70 min)

Retention time reproducibility for each peptide (n=6)

Peak No.

1

2

3

4

5

Retention Time(min)

49.08

54.48

57.95

61.26

63.93

m/z

511.63

583.9

653.34

526.28

582.32

Retention Time RSD(%)

0.64

0.63

0.56

0.54

0.55

Peak No.

6

7

8

9

10

Retention Time(min)

64.5

66.66

69.55

71.25

72.73

m/z

473.94

507.84

740.36

820.1

746.35

Retention Time RSD(%)

0.47

0.46

0.41

0.35

0.36

*(Reference)

When measuring the same sample by using the conventional LC system at 100 ¼LL/min, the retention time was 0.1% RSD.

Contact us for more information. For products and services outside North America, please click here.

 
 
starting of secondary navigation
Liquid Chromatography Systems

LaChromUltra - Ultra High-Speed LC

LaChrom Elite HPLC

Nano Flow LC System

LC/MS Systems

NanoFrontier eLD LIT-q-TOF LCMS System w/ECD

NanoFrontier LD LIT-TOF LCMS System

Amino Acid Analyzer

Spectrophotometers

UV-Visible/NIR Spectrophotometers

Fluorescence Spectrophotometers

Diffraction Gratings

starting of footer  | Site Map  | Terms of Use  |  Contact  |  Careers |

© 2002- Hitachi High Technologies America, Inc. All rights reserved.